Function "Full Calibr." (006); Display (008); Display - Pepperl+Fuchs Pulscon LTC Description Of Instrument Functions

Guided level radar with hart/4 ma ... 20 ma and profibus pa
Table of Contents

Advertisement

Pulscon LTC
Function group "basic setup" (00)
+
E
!
+
E
18
3.11

Function "full calibr." (006)

This function is used to enter the distance from the minimum level to the maximum level
(= span).
F
Note!
The usable measuring range lies between the lower and the upper blocking distance.
The values for empty distance (E) and span (F) can be set independently of this.
3.12

Display (008)

The distance measured from the reference point to the product surface and the
meas. value calculated with the aid of the empty adjustment are displayed. Check
whether the values correspond to the actual meas. value or the actual distance. The
following cases can occur:
Distance correct – meas. value correct  continue with the next function
"check distance" (051).
Distance correct – meas. value incorrect  check "empty calibr." (005)
Distance incorrect – meas. value incorrect  continue with the next function
"check distance" (051).
100 %
F = full calibration (= span)
F
= E – upper blocking distance
max
0 %