Differing from this, the following measuring error is present in the vicinity of the probe
end:
Rod probe and coax probe
80
60
40
20
0
-20
-40
-60
-80
0
50
100
150
Distance from probe end [mm]
Note!
Please re-enter the blocking distance in the function group "extended calibr." (05)
function "upper block. dist" (059) when installing the device in a high nozzle: upper
blocking distance (UB) = nozzle high (H) + 50 mm.
80
60
DC = 2
40
DC > 7
20
-20
-40
-60
-80
200
250
300
H
50
After 3 s, the following message appears
Rope probe
DC > 7
0
0
50
100
150
200
250
Distance from probe end [mm]
UB
300