Function group "output" (06), - "profibus param." (06), PROFIBUS PA only
8.11
–
+
E
If necessary, linearisation, the output signal and the current output can be tested with
the simulation function. You have the following simulation options:
Selection:
•
sim. off
•
sim. level
•
sim. volume
•
sim. current (HART only)
sim. off
Simulation is switched off.
sim. level
Enter the level value in "simulation value" (066).
The functions
•
measured value (000)
•
measured level (0A6)
•
output current (067) - only with HART instruments!
follow the entered values.
sim. volume
Enter the volume value in "simulation value" (066).
The functions
•
measured value (000)
•
output current (067) - only with HART instruments!
follow the entered values.
sim. current (HART only)
Enter the current value in "simulation value" (066).
The function
•
output current (067) - only with HART instruments!
follows the entered values.
Function "simulation" (065)
–
+
E
sim. volume
sim. level
sim. current
(HART only)
49