Pulscon LTC
Function group "length adjustment" (03)
5
–
+
E
5.1
–
+
E
Use this function to select the polarity of the probe end signal. If the probe end is
uncovered or in an insulated attachment, there is a negative probe end signal.
The signal from the probe end is positive if the attachment is grounded.
Selection:
•
free
•
tie down isol.
•
tie down gnd.
5.2
–
+
E
Use this function to select whether the probe length was changed after factory
calibration. Only then is it necessary to enter or correct the probe length.
Selection:
•
modified
•
not modified
Note!
!
If "modified" was selected in the "probe length" (031) function, the probe length is
defined in the next step.
5.3
–
+
E
Use this function to select whether the probe is at the time of the commisioning
uncovered or covered. If the probe is uncovered, the Pulscon LTC can determine the
probe length automatically "determine length" (034). function. If the probe is covered,
a correct entry is required in the "probe length" (033) function
Selection:
•
free
•
covered
Function group "length adjustment" (03)
Function "end of probe" (030)
Function "probe length" (031)
Function "probe" (032)
29