10 Initial commissioning

DANGER
The supply voltage and supply frequency must be the same as
given on the type plate.
Only connect the sterilizer to the specified socket.
CAUTION
Only use deionized or distilled water when topping up feed
water. Do not use tap water!
CAUTION
During the heating phase, contaminated exhaust air or waste
water may be emitted from the sterilizer to the on-site disposal
system.
Remedy: Expansion option Type FA exhaust air filtration.
INFORMATION
The central lock with thermolock is electro-mechanically
locked during delivery.
Connect to mains to unlock.
INFORMATION
The removal pressure the sterilizer is set to must be
assimilated to the ambient pressure of the installation site.
(See Chapter 9.)
If the selected removal pressure is too high, steam may escape
out of the lid at program end.
If the selected removal pressure is too low, the necessary
criteria for program end can not be achieved.
INFORMATION
Only use the condensate collector, if you just utilize the
sterilizer occasionally. When the collector overflows, water
damage may arise.
Especially in case of contaminated exhaust air, the water in the
collector as well as the collector itself may be infected with
germs. This factor is to be considered when disposing of the
collector content.
The steam sterilizer is marked with the name of the manufacturing company and a
type plate. All operating parts, such as valves, etc., are marked. A spare-parts
store for this instrument is maintained at the manufacturer's factory. Availability of
parts is guaranteed for 10 years.