6 Programs

The program keys are factory preset for complete program cycles. This means that
in addition to the process itself, the sterilizing time and temperature and certain
options for a sterilizing article are predefined. If you frequently sterilize the same
articles, this will be very comfortable.
Program
Base model:
Instruments
Destruction
(Waste)
Liquids
With Type RM water recooling system option only:
Liquids RM
With media temperature control expansion option only:
Liquids /
Liquids RM /
Liquids RO /
Liquids RG
With Type FA exhaust air filtration (
All programs
Table 2: Sterilizing programs
When you select a program, an indicator light in the corner of the selected program
key (18) is lit. Sterilizing temperature, multiple sterilization number (special
program), sterilizing time and program name appear in the cycle display (Figure 6).
121 C
Instruments
Figure 6: Cycle display
03/2007
Sterilization
121 °C / 20 min
134 °C / 15 min
121 °C / 20 min
121 °C / 20 min
121 °C / 20 min
Sterilizing time
actuated via media
temperature
Deaeration through
sterile air filter
Cycle
°
20min
6 Programs
Cooling
Fast steam release
Cyclical steam release
Cyclical steam release up
to 100 °C; program end at
90 °C reference
temperature
Water jacket cooling,
program end at 90 °C
reference temperature
Depending on program
(see above), program end
4)
at 90 °C
media
temperature
) only:
Cyclical steam release via
sterile air filter
26
DA0621_a/