Please observe the Commissioning and Operating Manual "Condensate Collector"
. There you find detailed information regarding the required connections,
maintenance information, important safety warning and much more.
If you wish to use your own container, please note the following:
The plastic container (e.g. PEHD) must be temperature-resistant and
able to contain approximately 20-30 litres.
The lid opening of the container must be large enough to facilitate easy
internal cleaning.
Automatic spent-steam condenser, water-cooled)
Disposing of the condensate:
By gully, siphon of washbasin or sink, waste-water line or condensate collector
(makeshift or used in safety laboratories for collecting and subsequently
disinfecting or sterilizing the condensate).
Disposing of tap water:
Heated cooling water does not come into contact with the condensate. Connection
to on-site cooling circuit is possible.
Cooling water: approximately 5 to 10 litres per batch.
Please observe our Installation and Maintenance Manual "Automatic spent-steam
condenser, water-cooled" . They include detailed information regarding the
required connections, maintenance information and important safety warnings.
Integrated spent-steam condenser with condensate return
When on-site disposal systems are completely missing or when operated in safety
laboratories, the sterilizer may be equipped with a condensate return system.
Condensate produced during sterilization is returned to the feed water tank,
resulting in a closed circuit system.

10.4.2 Connecting the steam sterilizer to the supply system:

Type
25T
Table 4: Power supply requirements on the building site
The power cable is located at the back and underneath the sterilizer. The mains
socket must be located in the immediate vicinity of the sterilizer. The mains cable is
2.5 m long.
♦ Plug the power cable into a socket outlet.
03/2007
10 Initial commissioning
1
1
Figure 16: Condensate collector
Heating
Mains
System
connection
3.0 kW
230 VAC / 50 Hz
47
Quick-release
coupling
Socket / Fuse
Socket outlet with grounding
contact 16 AT
DA0621_a/

Hide quick links: