Initial Operation; Initial Sanitization - Thermo Scientific EASYpure 1164 Series Operation Manual And Parts List

Table of Contents

Advertisement

Warning
Avoid splashing disinfecting solutions
on clothing or skin.
Ensure all piping connections are tight
to avoid chemical leakage.
Ensure adequate ventilation.
Carefully follow manufacturerʼs safety
instructions on labels of chemical con-
tainers and material safety data
sheets.
This device is to be used with water
feeds only.
Sanitizing/cleaning agents must be
used in compliance with instructions in
this manual. Failure to comply with the
above could result in explosion and
personal injury.
Use of a properly grounded electrical
outlet of correct voltage and current
handling capacity.
Figure 9: Preparing To Sanitize

Initial Operation

Initial Sanitization

Your EASYpure UV/UF has been shipped with a sanitiza-
tion cartridge (Catalog Number D50245) and two empty
cartridges to allow you to sanitize your EASYpure UV/UF
prior to using it for the first time. To sanitize your
EASYpure UV/UF:
Sanitization Cartridge
D50245
Empty Cartridges
1.
Open the cartridge access door. Install the
supplied D50245 sanitization cartridge in the
empty cartridge position. Press the upper end
cap into the upper right position until it bot-
toms out. Then, lower the cartridge and insert
the lower end cap into the lower socket until it
is firmly seated. The empty cartridges have
been installed for you at the factory. Close
and latch the cartridge access door.
2.
Insert the quick disconnect insert on the feed-
water line into the feedwater inlet quick dis-
connect on the back of the EASYpure UV/UF.
3.
Plug the power cord into the unitʼs power
entry module and plug into a live outlet. Place
the main power switch to the on position.
Place a suitable container under the outlet
valve. Depress the draw-off lever to open the
outlet valve. Press the "Flush" button twice to
begin an extended flush.
4.
Close the draw-off valve when solution exits
the system.
5.
When the extended flush stops (5 minutes),
open the draw-off valve. Press the
START/STOP button to start. Draw off 2.0
liters of solution through the draw-off valve.
Close the draw-off valve. Discard the solution.
6.
Recirculate the disinfectant solution for ap-
proximately 30-45 minutes.
7.
Press the START/STOP button to stop.
17