♦ Open the door a crack. When doing so, hold the door by the lower third of
its right edge. Do not put your fingers behind the door.
DANGER
Hot steam may escape when the door is opened.
Scalding hazard!
Step back when opening the sterilizer.
♦ Wait one minute.
♦ Now, rotate the door to the side.
DANGER
After drying in the sterile air stream, the sterilizing chamber
and the sterilized items will be very hot!
INFORMATION
Without the use of additional drying equipment, filters, fabric,
or cotton wool will not dry properly after sterilization.
INFORMATION
After sterilization while using a flexible temperature sensor:
Remove the flexible temperature sensor from the reference
flask of leave the reference flask in the sterilizing chamber (8).
♦ Remove sterilized items.
The sterilizer is ready for a new cycle immediately after the sterilized items have
been removed and the feed water tank has been filled up (if required) with
deionized water.
INFORMATION
The door (2) should always be left open when the sterilizer is
not in use so that remaining moisture may disappear.
03/2007
8 Program end and removal
37