Thermo Scientific VARIOKLAV 25T Operating Manual page 56

Steam sterilizer
Table of Contents

Advertisement

Message in the
Meaning
cycle display
Error 17
Heating system does not turn off.
Overheating
Heating system relay defective or
program control error.
Error 20
Microheating system (heatable
Microheating
temperature sensor) cannot regulate
water feed to jacket.
Microheating system defective.
Error 22
Heating coils are defective, or
No heater
the heating system control is
defective.
Error 23
One heating coil is defective.
Partial heater
Err1
Sensor broken
in the Chamber
Chamber temperature sensor is
field:
affected.
Flexible temperature sensor
in the Media field:
(reference temperature sensor) is
affected
Err2
Sensor shorted
in the Chamber
Chamber temperature sensor
field:
Flexible temperature sensor
in the Media field:
(reference temperature sensor).
Err3
Pressure sensor defective
in the Chamber
Pressure sensor in sterilizing
field:
chamber
in the Jacket field:
Pressure sensor in jacket
below the Jacket
Pressure sensor in seal
field:
Table 5: Error codes and their meanings
12)
If present
03/2007
12 Troubleshooting
12
.
12)2)2)
12)
12)
56
Remedy
Contact after-sales
service.
Contact after-sales
service.
Contact after-sales
service.
Contact after-sales
service.
Have technician
replace sensor or
contact after-sales
service.
Have technician
replace sensor or
contact after-sales
service.
Contact after-sales
service.
DA0621_a/