7 Starting and running a program
In the case of the Multiple sterilization special program, the program cycle of
heating, sterilizing, cooling and drying (optional) repeats itself according to the
selected multiple sterilization number.
In the Liquids programs, the thermolock will unlock the door (2) as soon as the
reference temperature drops below 90 °C.
By pressing the START/STOP key (19), you can interrupt the program cycle at any
time. The sterilizer always returns to a safe state before unlocking the door.
If the program is followed by a drying phase, the current values display (13) shows
the drying time and the ▼ symbol. The drying time counts down to zero.
33
03/2007
DA0621_a/