11 Maintenance and cleaning

DANGER
Repairs and interventions may only be carried out by our
authorized after-sales service.
DANGER
If the safety valve does not react to a higher pressure than the
set nominal pressure, cease using the sterilizer immediately.
Do not use it again until a new and fully functional safety valve
is installed.
DANGER
When a fault occurs and for maintenance purposes,
disconnect the steam sterilizer from the mains supply.
Pull the mains plug or neutralise the fuse.
INFORMATION
When introducing "Deodorant Pearls" into the sterilizing
chamber (8) in order to avoid unpleasant odours:
Some ingredients of the "Deodorant Pearls" may rapidly
disintegrate the door seal (6)
INFORMATION
Only original seals made by the manufacturer may be used.
INFORMATION
We highly recommend you to have the sterilizer serviced
regularly by our authorized after-sales service.
We also recommend to enter into a contract for maintenance
or for annual safety checks.
Repeat testing of the pressure vessel must only be carried out
by an authorized person fulfilling the requirements stated in
the German Operating Safety Regulation.
03/2007
11 Maintenance and cleaning
51
DA0621_a/