Thermo Scientific VARIOKLAV 25T Operating Manual page 16

Steam sterilizer
Table of Contents

Advertisement

21 Acoustic signal key
22 Preheating key
23 Timer key
24 Sterilization phase
display
25 Printer cover
26 Printer ON/OFF key
27 Paper feed key
The VARIOKLAV Type 25 T steam sterilizer is a compact benchtop unit with an
electrical heater built into the jacket space:
The feed water is pumped from the feed water tank to the jacket space where it is
heated. The heated feed water then supplies the sterilizing steam. The sterilizing
steam is introduced directly into the sterilizing chamber (8). A rigidly mounted
temperature sensor is measuring the chamber temperature.
The following components are stored in the case (3):
the sterilizing chamber (8),
the safety valve,
the feed water tank (4),
the control panel (5).
The control panel (5) is used to select sterilization parameters and to start the
sterilizing program. Sterilizing chamber (8) pressure and temperature are shown
on the control panel display.
The built-in timer is used to set a start time for the sterilization cycle.
The sterilizer may include additional expansion options indicated on the front page.
Expansion options are described in Chapter 4.
We highly recommend having the sterilizer serviced regularly by an authorized
service agency. Furthermore, please observe the Maintenance Manual. This will
give you guaranteed trouble-free, safe operation and a long service life.
We offer you a low-priced yearly safety check and benefits such as calibration and
qualification (see Chapter 10.6).
03/2007
3 Equipment description
: Turns the acoustic signal (buzzer) at program
end on or off.
: Out of function
: Turns the timer on or off, and for checking the
preset date or time.
: Shows the phases of a sterilization cycle.
: Covers the printer head, the ribbon, and the
paper roll (option).
: Turns the printer on or off.
: Advances the printer paper.
16