Batch Documentation (Option) - Thermo Scientific VARIOKLAV 25T Operating Manual

Steam sterilizer
Table of Contents

Advertisement

7.1

Batch documentation (option)

Integrated printer
The integrated printer (if present) logs sterilization process data. The printout is
tamper-proof. The log header (1) is printed out as soon as the program is started.
Chamber temperature, media temperature and chamber pressure are continuously
measured during program execution. The log shows the measured values of the
sterilization phase in increments of one minute. Values are logged every 10
minutes during the heating and cooling phases.
Errors occurring during sterilization are printed immediately.
= = = = = = = = = = = = = = = = = = = = = = =
- - - - - - - - - - - - - - - - - - - - - - -
D AT E
1
T I M E
C h a r g e N o
- - - - - - - - - - - - - - - - - - - - - - -
I n s t r u m e n t s
1 2 1 ° C
a ) C h a m b e r Te m p [ ° C ]
b ) L i q u i d Te m p [ ° C ]
c ) C h a m b e r P r e s s [ k P a ]
- - - - - - - - - - - - - - - - - - - - - - -
H E AT I N G
2 1 : 0 2
2 1 : 1 2
2 1 : 2 2 1 0 5 . 5 1 2 0 . 9 1 5 0
S T E R I L I Z I N G
2 1 : 3 1 1 2 1 . 2 1 2 2 . 3 3 2 0
2 1 : 3 2 1 2 1 . 2 1 2 2 . 9 3 2 0
2 1 : 3 3 1 2 1 . 2 1 2 2 . 9 3 2 0
2 1 : 3 4 1 2 1 . 2 1 2 2 . 3 3 2 0
2 1 : 3 5 1 2 1 . 2 1 2 2 . 9 3 2 0
2 1 : 3 6 1 2 1 . 1 1 2 2 . 9 3 2 0
2
2 1 : 3 7 1 2 1 . 2 1 2 2 . 3 3 2 0
2 1 : 3 8 1 2 1 . 2 1 2 2 . 9 3 2 0
2 1 : 3 9 1 2 1 . 3 1 2 2 . 9 3 2 0
2 1 : 4 0 1 2 1 . 2 1 2 2 . 3 3 2 0
2 1 : 4 1 1 2 1 . 2 1 2 2 . 9 3 2 0
2 1 : 4 2 1 2 1 . 2 1 2 2 . 9 3 2 0
2 1 : 4 3 1 2 1 . 2 1 2 2 . 3 3 2 0
2 1 : 4 4 1 2 1 . 2 1 2 2 . 9 3 2 0
2 1 : 4 5 1 2 1 . 2 1 2 2 . 9 3 2 0
2 1 : 4 6 1 2 1 . 0 1 2 2 . 9 3 2 0
2 1 : 4 7 1 2 1 . 2 1 2 2 . 3 3 2 0
2 1 : 4 8 1 2 1 . 2 1 2 2 . 9 3 2 0
2 1 : 4 9 1 2 1 . 3 1 2 2 . 9 3 2 0
2 1 : 5 0 1 2 1 . 3 1 2 2 . 3 3 2 0
2 1 : 5 1 1 2 1 . 2 1 2 2 . 9 3 2 0
R E C O O L I N G
2 1 : 5 2 1 2 1 . 1 1 2 2 . 9 3 2 0
2 2 : 0 2
2 2 : 1 2
2 2 : 2 2
E N D O F C Y C L E
2 2 : 2 3
- - - - - - - - - - - - - - - - - - - - - - -
S t e r i l i z i n g C y c l e s u c c e s s f u l
3
S I G N .
= = = = = = = = = = = = = = = = = = = = = = =
Figure 12: Sterilization log (example)
03/2007
7 Starting and running a program
VA R I O K L AV 2 5 T
0 0 1 2 3 4
2 0 . 0 3 . 0 7
2 1 : 0 2
1 7
2 0 m i n
a )
b )
c )
6 2 . 8
6 0 . 2 1 0 0
9 5 . 2
9 6 . 9 1 0 0
9 8 . 2
9 9 . 7 1 0 0
7 3 . 9
7 5 . 8 1 0 0
6 4 . 3
6 4 . 4 1 6 0
5 6 . 3
6 4 . 4 1 0 0
Log
: Device ID and batch
1
header
information (date, time,
batch number) and
program parameters.
2
Values
: Temperature and
pressure values for the
various program phases
at the time indicated
(first column)
Log
: Successful sterilization
3
footer
notice, signature field.
34