6-6 Calibration Settings
Parameter name
Calibration select (CAL.SEL)
(1) Measure zero
Target point
Zero point
Operation procedure
■
Long press
for 2 s
Note
■
Calibration cannot be executed in the following cases. If they occur, an ERROR is shown on digital
display MAIN (white).
The span and measurement values have different signs. (The target value and detected value
are opposite in sign.)
Relative to the span at time of shipment, span adjustment gives an adjustment ratio greater
than two.
The measurement value is indefinite. ("------" is displayed.)
The measurement value is out of measurement range. (+OVER or −OVER is displayed.)
When a sensor head is replaced, carrying out zero and span
adjustment reduces the errors existing at installation time.
DEFAUL (Default) / USER (Set by user)
(2) Measure target
point.
point.
Select
Commit
<(1) Register zero point.>
<(2) Register target point>
<(3) Input theoretical value>
6- 50
Setting value
[Default: DEFAUL]
(3) Enter theoretical value of span
using numeric input.
Span
Cancel
Select
Commit
Commit
Commit
Automatic
Remarks
Shift between
Change numerical
digits