6-1-3 Hysteresis (HYSTER)

Parameter name
Hysteresis (HYSTER)
With HYSTER set to 0:
Chattering makes judgment output unstable.
HIGH setting
value
HIGH設定値
LOW setting
value
LOW設定値
HIGH
GO
LOW
Operation procedure
Long press
2秒長押し
for 2 s
HI.SET
LO.SET
Select
選択
Commit
確定
Note
HIGH and LOW share the same hysteresis value.
Set hysteresis to a value smaller than the difference between the HIGH and LOW setting values. If
hysteresis is set to a value equal to or exceeding the difference between the HIGH and LOW
setting values, the GO judgment will not be output.
Setting value
Judgment may be unstable due to measurements fluctuating
around the HIGH or LOW setting value. A large hysteresis value
can prevent this chattering effect.
0.0000 to 199.9999 (mm)
With HYSTER set to non 0:
Judgment output becomes stable.
HIGH setting
HIGH設定値
LOW setting
LOW設定値
HIGH
LOW
ADVANC
CALC
COPY
BANK
CALIB
/
TEACH
TOL <±>
PR.VAL
Cancel
キャンセル
Shift between
Change numerical
digits
values
桁移動 数値変更
6- 7
[Default: 0.0030]