Citizen SA-SD Operation Manual page 70

Controller for contact-type displacement sensors
Table of Contents

Advertisement

6-2-8-1 Alarm Delay Time
Parameter name
Alarm delay time (DELAY)
Can be set such that a momentary upthrust will not activate the alarm.
Operation procedure
Long press
for 2 s
6-2-8-2 Upthrust Check
Parameter name
Upthrust check
(PRS.CHK)
Operation procedure
Long press
for 2 s
Setting value
Sets the delay time (changeable in 1 ms increments) from the
time the sensor head comes into upthrust position to the time the
alarm activates.
1 to 1,000
Select
Commit
Setting value
Sets the upthrust check ON/OFF.
ON / OFF
Select
Commit
6- 40
Remarks
[Default: 1,000]
Cancel
Shift between
Change numerical
digits
values
Remarks
[Default: ON]
Cancel