Alarm Settings - Citizen SA-SD Operation Manual

Controller for contact-type displacement sensors
Table of Contents

Advertisement

6-2-8 Alarm Settings

Parameter name
Alarm delay time (DELAY)
Upthrust check
(PRS.CHK)
Upthrust check setting value
(PRS.SET)
Catching check
(CAT.CHK)
Number of connected units check
(PRS.CHK)
Upthrust
"Upthrust" indicates a state in which the spindle has been pushed in too much. Upthrust can cause damage to
the sensor head. If the upthrust check is set to ON, an alarm is output before upthrust occurs, protecting the
sensor head from damage.
Setting value
Sets the delay time (changeable in 1 ms increments) from the
time the sensor head comes into upthrust position to the time the
alarm activates.
1 to 1,000
Sets the upthrust check ON/OFF.
ON / OFF
Sets the threshold level of the upthrust check.
−199.9999 to 199.9999 (mm)
Sets the catching check ON/OFF.
ON / OFF
If controllers are connected, this performs a check when the
power is turned ON confirming that the number of connected units
has not changed.
ON / OFF
6- 39
Remarks
[Default: 1,000]
[Default: ON]
[Default: 10.0000]
[Default: OFF]
[Default: OFF]

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents