5-7 Preset

Preset is a function that determines a reference for measurement values using a master work, etc.
When the preset operation is performed, the current value becomes the preset value.
If 0 is set as the preset value, the current value becomes 0.
Press the Preset key while a master work is being measured.
The preset function comes ON, and the preset indicator lamp (green) lights up.
To quit the preset function, hold the Preset key down for two seconds.
Short press
For reference
If preset is performed using keys, the preset data is retained after the power is turned off because
it is written into internal memory. (Values can be overwritten approximately one million times.)
Preset cannot execute correctly in the following cases:
Immediately after the power is turned on, or immediately after input of a reset
When the data is outside the display's upper or lower limit.
When an alarm is being output
When an error is being output
Related information
For details of how to set preset values, refer to "6-1-6 Preset (PR.VAL)."
If the preset input is set to external input, the external input can be used to perform the preset.
For details of preset input from external input, refer to "6-2-3 External Input."
<Preset>
5- 8
<Quit preset>
Long press for 2 s