Low Setting Value (Lo.set) - Citizen SA-SD Operation Manual

Controller for contact-type displacement sensors
Table of Contents

Advertisement

6-1-2 LOW Setting Value (LO.SET)

Parameter name
LOW setting value (LO.SET)
Operation procedure
Long press
2秒長押し
for 2 s
HI.SET
Select
選択
Commit
確定
Note
Configure settings such that HIGH setting value > LOW setting value.
Although it is possible to set HIGH setting value < LOW setting value, the GO judgment will not be
output.
Related information
The LOW setting value can always be displayed on digital display SUB (green).
For details, refer to "5-2 SUB Display Switching Mode."
HIGH and LOW setting values can be set by teaching using a master work, upper limit work, and
lower limit work.
For details, refer to "5-3 Teaching Mode."
Setting value
Sets a LOW setting value for pass/fail judgment.
−199.9999 to 199.9999 (mm)
: Included
: Not included
HIGH
GO
LOW
A DV A N C
A DV A N C
A DV A N C
A DV A N C
CALC
COPY
BANK
/
HYSTER
TEACH
TOL <±>
Cancel
キャンセル
Shift between
Change numerical
digits
values
桁移動 数値変更
6- 6
[Default: 1.0000]
HIGH setting value
LOW setting value
CALIB
RESET
MAINTE
PR.VAL
PR.OBJ
PR.SAVE
DIRECT
LEVER
Remarks