6-2-8-4 Catching Check
Parameter name
Catching check
(CAT.CHK)
If, due to aging, spindle movement becomes poor and the spindle does not return to the correct position, then
accurate measurement cannot be performed. The catching check function detects failure of the spindle to
return correctly.
This function checks for complete extension of the spindle while the trigger input is ON by detecting whether
the normal measurement value (NORM.V) falls below the LOW setting value.
If the normal measurement value has not fallen below the LOW setting value whilst the trigger input is ON, an
alarm is activated.
The alarm can also be set not to activate.
HIGH setting value
LOW setting value
TRIG
Judgment output
Operation procedure
Long press
for 2 s
Note
This function can be used only when the measurement mode is set such that external trigger input
has an effect, as described in "6-2-1 Hold Settings."
Setting value
Sets the catching check ON/OFF.
ON / OFF
Alarm
Select
Commit
6- 42
Remarks
[Default: OFF]
Cancel