3.7.3 Digital Output C (cont.)
3.7.3c
Setting the Frequency (cont.)
Example 1:
Base: 0 m/s = 0 Hz
Full: 10 m/s = 100 Hz
Then y Hz = x (m/s)
10 Hz
Example 2
Base: 10 kg = 0 Hz
Full: 20 kg = 10 Hz
Then y Hz =(x kg -10)
PanaFlow HT User's Manual
1000
To check the current frequency value, on the Frequency display select Frequency
Value and press
To change the current value, press
page 59.
The Base Value is the measurement value represented by 0 Hz. To check the
current base value, on the Frequency display select Base Value and press
screen similar to the following appears.
[
]
. A screen similar to the following appears.
and proceed as in Setting the Pulse on
[
]
Chapter 3. Initial Setup and Programming
[
]
. A