Appendix G. CE Mark Compliance
G.1 Introduction
For CE Mark compliance, the PanaFlow HT flowmeter must be wired in accordance with the instructions in this
appendix.
IMPORTANT:
CE Mark compliance is required only for units intended for use in EC countries.
G.2 Wiring
The PanaFlow HT must be wired with the recommended cable, and all connections must be properly shielded and
grounded. Refer to Table 15 for the specific requirements.
Connection
Transducer
Armored RG62 a/U
Input/Output
Armored 22 AWG shielded (e.g. Baystate #78-1197) with
armored material added to outside of jacket
Power
Armored 14 AWG 3 conductor
Note:
If the Panaflow HT is wired as described in this appendix, the unit will comply with the EMC Directive
2004/108/EC.
PanaFlow HT User's Manual
Table 15: Wiring Requirements
Cable Type
Appendix G. CE Mark Compliance
Ground Termination
Grounded using a cable gland.
Grounded using a cable gland.
Grounded using a cable gland.