4-4-10 PW Doppler Mode Checks
4-4-10-1 Introduction
4-4-10-2 Preparations
Versana Balance – Basic Service Manual
5808768-100 English Rev.9
PW Doppler are used to measure velocity (most often in blood).
Doppler mode can be done with a special pencil probe or with
an ordinary probe. By using an ordinary probe, you can first
bring up a 2D picture for navigation purpose and then add PW
Doppler.
•
Connect one of the probes to the scanner.
•
See
4-4-13 'Probe/Connectors Check' on page 4-66
about connecting the probes.
For available probes, see
•
Turn ON the scanner
The 2D Mode window is displayed (default mode).
•
If needed, adjust the Display's Brightness and Contrast
setting.
Press PW to start Pulsed Wave Doppler (PW).
Use the trackball to select the Area of Interest (Sample
Volume) in PW.
Functional checks
9-3-2 'Probe' on page
System view
for info
9-6:
4-61