Electrical Safety Checks - GE MAC 800 Service Manual

Resting ecg analysis system
Hide thumbs Also See for MAC 800:
Table of Contents

Advertisement

Electrical Safety Checks

Step
Earth Leakage Current
1
Forward Polarity
2
Neutral Open, Forward Polarity
3
Neutral Open, Reverse Polarity
4
Reverse Polarity
Enclosure Leakage Current
1
Forward Polarity
2
Neutral Open, Forward Polarity
3
Ground Open, Forward Polarity
4
Ground Open, Reverse Polarity
5
Neutral Open, Reverse Polarity
6
Reverse Polarity
Patient Leakage Current To Ground
1
Forward Polarity
2
Neutral Open, Forward Polarity
3
Ground Open, Forward Polarity
4
Ground Open, Reverse Polarity
2031504-159D
11.
Verify the battery test was successful.
See
"Battery Test"
on page 3-11 for more information.
12.
Verify the LAN test was successful.
See
"LAN Test"
on page 3-14 for more information.
13.
Verify the modem test was successful.
See
"Modem Test"
on page 3-15 for more information.
14.
Verify the current leakage test results meet the requirements.
Perform electrical safety checks when indicated. All indicated electrical safety
checks require a pass/fail indication for the steps performed. Record the
measurement values in your debrief.
Electrical Safety Checks
1
Condition
UUT - ON
NC
______
SFC
______
SFC
______
NC
______
NC
______
SFC
______
SFC
______
SFC
______
SFC
______
NC
______
NC
______
SFC
______
SFC
______
SFC
______
MAC™ 800
2
Result

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail

Pass/Fail
Maintenance
Leakage Current Limits

500

1,000

1,000

500

100

500

500

500

500

100

10

50

50

50
4-27