Do you have a question about the Sara 3000 and is the answer not in the manual?
Questions and answers
Savannah
April 28, 2025
Just bought a used sara 3000. Previous owner said it needed a new battery. Checked the battery and it works and charges fine. However when the battery is put in place and connected to the machine- the machine does not turn on. There is no response what so ever. No lights or sounds, completely unresponsive. Checked underneath by the switches and wiring. Nothing visibly wrong with it. All wires look new, nothing disconnected. Tried unscrewing the connector plug for the batter to the machine thinking maybe its not being pushed in fully for the battery to connect. Tried that and still nothing. I need this Machine for my grandfather and could not afford a brand new machine. Please any information helps.
1 comments:
Mr. Anderson
May 15, 2025
If the Arjohuntleigh Sara 3000 is not turning on despite a functioning battery, one possible reason is that the red Power Off button is still engaged. The resolution is to press the green Power On/reset button.
Sara 3000 Instructions for Use KKX81010M_EN_12 • 04/2016 ...with people in mind...
Page 2
This product has been manufactured for ArjoHuntleigh by: ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö, Sweden ‘Sara 3000’, ArjoHuntleigh’ and ‘ARJO’ are registered trademarks of GETINGE AB. The Sara 3000 is produced in Poland by ArjoHuntleigh AB for GETINGE AB and sold under the ArjoHuntleigh brand.
Safety Instructions Intended use General Safety Instructions Sara 3000 is a mobile raising aid, with a Safe Any references made to the ‘resident’ or ‘patient’ in Working Load of 200 kg (440 lbs), intended to be this document, means the person being raised.
Safety Instructions Symbols and Definitions used in this Product Lifetime document The expected operational life of the Sara 3000 is ten (10) years, providing the product has been regularly Attention - special disposal requirements serviced and maintained as indicated in these instructions (see the “Care and Preventive...
Any abridged guidelines are not a replacement of the full instructions for Sara 3000, its parts and ArjoHuntleigh sling. Before operating Sara 3000 be sure that you read, understand and follow all the instructions and guidelines for both Sara 3000 and the ArjoHuntleigh sling.
Parts designation / legend Parts referred to in this manual Control Panel Transfer Sling Standing Sling Fig. 1 Clip Attachment Points 14. Chassis Vertical Mast Battery Pack 15. Scale Unit Display (if fitted) Steering Handgrips 16. Front Castors Hand Control 17.
On the top control panel. Press this button to turn on power to the Sara 3000. Also used to reset if the automatic overload fuse has operated (indicated by Features and functions the button projecting outwards slightly).
15 seconds and “OverHeat” is displayed on castors have brakes which can be foot operated if the Hour/Cycle Meter (see Fig. 1) when the required (see Fig. 4), to keep the Sara 3000 in operator exceeds the duty cycle for the lift actuator position.
Although a resident can be put into the sling with the Sara 3000 close by, it may be easier to do so with the Sara 3000 moved away. For the transfer of residents and use of various slings refer to the instruction provided by the manufacturer of the sling.
Page 12
(see Fig. 5). The strap also retains the sling in the correct position around the resident. If the Sara 3000 is not already close to the resident, bring it to the resident. Apply the clips to the attachment points, and secure them for safety by pulling the clips as shown in Fig.
Page 13
If required, the chassis legs may be opened to go around a chair, by operating the appropriate control button. Carefully push the Sara 3000 in closer to make full lower leg contact with the knee support. (see Fig. 9) Fig. 10 When raising the resident with a standing sling, the resident’s body posture shall go from seated to...
Page 14
Fig. 11 Warning: Always move the Sara 3000 When using the standing sling, the resident can be in the direction shown in Fig. 11. raised to a fully standing position (see Fig. 12).
Place the battery pack on charge as follows: To ensure the Sara 3000 is always ready for use, it is recommended that a freshly charged battery pack When the battery discharge indicator displays 3 is always available.
Page 16
Sara 3000. An electrical connection is made automatically. Ensure the green Power on/Reset button is pressed The Sara 3000 is now ready for use. Caution: After use, turn off the Sara 3000 by pressing the red Power off button (Fig.
MRSA and several other micro-organisms under light soiling conditions. Warning: ArjoHuntleigh recommends Warning: The lifter should be cleaned that Sara 3000 be maintained at regular before it is used by another patient. intervals. See Preventive Maintenance Schedule section in this document.
Disinfection, Cleaning and Maintenance Parts lists and circuit diagrams Environmental Advice These are available on request from ArjoHuntleigh This device is marked with the WEEE or its approved distributors. symbol (crossed-out wheeled bin) to indicate that it is electronic equipment...
Care and Preventive Maintenance The Sara 3000 is subject to wear and tear, and the following actions must be performed when specified to ensure that the product remains within its original manufacturing specification. Warning The points on this checklist are the minimum the manufacturer recommends. In some cases due to heavy use of the product and exposure to aggressive environment, more frequent inspections shall be carried out.
58.2 Sara 3000 with Scale (without Battery Pack) 67.9 The Sara 3000 is not intended to be broken up into more than these parts. Safe Working Load Sara 3000 All slings - please verify with the SWL on the sling label...
Technical Specification Conforms to UL standard 2601-1 and certified to CAN/CSA standard C22.2 No 601.1-M90 Duty cycle/Max volts/Max amps Mast Lift Actuator 10% - (2 min/18min)/24V/10,5A ”V” Chassis Actuator 10% - (2 min/18min)/24V/2,2A Mode of operation: intermittent Environment Air Humidity/Storage 80% at 20°C (68°F) Usage temperature range (ambient) +5°C (41°F) to +35°C (95°F)
‘C.S.P.’ stands for ‘Central Suspension Point’: a reference point on the Sara 3000 used for measurements. As CSP on the Sara 3000 we have used the clips attachment point closest to the resident at the start of the raising cycle.
This will indicate the power level of the battery. If in doubt, replace the battery on the Sara 3000 with a fully charged one and compare performance. In case of low battery power level, charge the battery and use a fully charged one to continue using the Sara 3000.
RF communications equipment (transmitters) and the Sara 3000 by maintaining a distance of not less than 1 metre to any part of the Sara 3000, including cables, if the transmitter’s output power rating exceeds 1W.
Page 27
Free: +1 800 072 040 Fax: +33 (0) 3 20 28 13 14 Fax: + 61 89337 9077 E-mail: info.france@arjohuntleigh.com POLSKA ArjoHuntleigh Polska Sp. z o.o. BELGIQUE / BELGIË HONG KONG ul. Ks Piotra Wawrzyniaka 2 ArjoHuntleigh NV/SA ArjoHuntleigh (Hong Kong) Ltd...
Page 28
Getinge Group is a leading global provider of products and systems that contribute to quality enhancement and cost ef ciency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, Getinge and Maquet. Getinge provides solutions for infection control within healthcare and contamination prevention within life sciences. Maquet specializes in solutions, therapies and products for surgical interventions, interventional cardiology and intensive care.
Need help?
Do you have a question about the Sara 3000 and is the answer not in the manual?
Questions and answers
Just bought a used sara 3000. Previous owner said it needed a new battery. Checked the battery and it works and charges fine. However when the battery is put in place and connected to the machine- the machine does not turn on. There is no response what so ever. No lights or sounds, completely unresponsive. Checked underneath by the switches and wiring. Nothing visibly wrong with it. All wires look new, nothing disconnected. Tried unscrewing the connector plug for the batter to the machine thinking maybe its not being pushed in fully for the battery to connect. Tried that and still nothing. I need this Machine for my grandfather and could not afford a brand new machine. Please any information helps.
If the Arjohuntleigh Sara 3000 is not turning on despite a functioning battery, one possible reason is that the red Power Off button is still engaged. The resolution is to press the green Power On/reset button.
This answer is automatically generated