Teledyne Foxy R1 Instruction Manual page 71

Hide thumbs Also See for Foxy R1:
Table of Contents

Advertisement

Foxy
NONPEAK=n
This command sets the non-peak/window
action of the current program. Choices are Collect (n=1), Divert
(n=2), or Drain (n=3).
Example:
Command:
Response:
The NONPEAK command without an operand returns the
current setting.
Example:
Command:
Response:
PATTERN=n
This command sets the rack filling pattern of
the current program. Choices are Standard (n=1) or Left-to-right
only (n=2).
Example:
Command:
Response:
The PATTERN command without an operand returns the current
setting.
Example:
Command:
Response:
PROGRAM=n
This command sets the current program number. Choices are n=1
to 8.
Example:
Command:
Response:
The PROGRAM command without an operand returns the
current setting.
Example:
Command:
Response:
®
®
R1 and Foxy
R2 Fraction Collectors
Section 4 Serial Command Control
NONPEAK=3
(none)
NONPEAK
NONPEAK=3
PATTERN=1
(none)
PATTERN
PATTERN=1
PROGRAM=8
(none)
PROGRAM
PROGRAM=8
4-11