Section 4 Serial Command Control
4-10
Example:
Command:
Response:
FSIZE=n
This command sets the fraction size for the current
program. If FTYPE=1 (fraction by time), n=1 to 359999 seconds; if
FTYPE=2 or 3 (fraction by drops or external counts), n=0 to
999999.
Example:
Command:
Response:
The FSIZE command without an operand returns the current
setting.
Example:
Command:
Response:
FTYPE=n
This command sets the fraction type of the current
program. Choices are Fraction by Time (FTYPE=1), Fraction by
Drops (FTYPE=2), or Fraction by External counts (FTYPE=3).
Example:
Command:
Response:
The FTYPE command without an operand returns the current
setting.
Example:
Command:
Response:
LAST=n
This command sets the last tube of the current
program. Choices are n=1 to n=maximum number of tubes for
programmed rack type.
Example:
Command:
Response:
The LAST command without an operand returns the current
setting.
Example:
Command:
Response:
DELAY
DELAY=102
FSIZE=23
(none)
FSIZE
FSIZE=23
FTYPE=3
(none)
FTYPE
FTYPE=3
LAST=10
(none)
LAST
LAST=10