CE Declaration of Conformity
DECLARATION OF CONFORMITY
Application of Council Directive:
Equipment Type/Environment:
Standards to which Conformity is Declared:
Standard
EN61000-3-2:2014
EN61000-3-3:2013
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-11
CISPR11/
EN 55011
The undersigned, hereby declares that the design of the equipment specified above conforms to the above Directive(s) and Standards
as of May 31, 2016.
USA Representative
Edward J. Carter
Director of Engineering
60-2132-663
Rev B
Manufacturer's Name:
Manufacturer's Address:
Trade Name/Model No:
Year of Issue:
Description
Current Frequency Harmonics
Voltage Fluctuations and Flicker
Electrostatic Discharge
Radiated RF Immunity
Electrical Fast Transient
Surge on AC Lines
Conducted RF on AC lines
Voltage Dips/Interruptions
RF Emissions
2014/30/EU -The EMC Directive
2002/96/EC – The WEEE Directive
2014/35/EU – The Low Voltage Directive
Teledyne Isco
4700 Superior
Lincoln, Nebraska 68504-1398 USA
P.O. Box 82531, Lincoln, NE 68501-2531
Phone:
+1 (402) 464-0231
Facsimile:
+1 (402) 465-3799
Laboratory Equipment for Light Industrial/Commercial Environments
Foxy R1, R2
2016
rd
EN 61010-1 3
edition
Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
EN 61326-1:2013
EMC Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
Severity Applied
Class A
Class A
Level 2 - 4kV contact discharge
Level 3 - 8kV air discharge
80 MHz to 1000MHz 80% AM at 1kHz
Level 2 - 3V/m
Level 2 - 1kV on AC lines
Level 2 - 1kV common mode,
Level 2 - 0.5KV differential mode
150 kHz to 80 MHz,
Level 1 - 1V rms, 80% modulated
100% drop, 10ms
Group 1, Class A Industrial, Scientific, and
Medical Equipment
Performance Criteria
PASS
PASS
A
A
A
A
A
A