Section 4 Serial Command Control
4-12
RANGE=n
This command sets the detector signal voltage range of the
current program. The choices are:
• n=1 1 Volt
• n=2 100 mV
• n=3 10 mV
• n=4 External detector
Example:
Command:
Response:
The RANGE command without an operand returns the current
setting.
Example:
Command:
Response:
RESTART=n
This command sets the restart action of the
current program.
Choices are:
• n=1 None
• n=2 Inject, Overlay
• n=3 Inject, Skip Tube
• n=4 Time, Overlay
• n=5 Time, Skip Tube
• n=6 Inject, Next Tube
• n=7 Time, Next Tube
Example:
Command:
Response:
The RESTART command without an operand returns the current
setting.
Example:
Command:
Response:
RSTIME=n
This command sets the restart time (n=0 to
359999 seconds) of the current program.
Example:
Command:
Response:
RANGE=2
(none)
RANGE
RANGE=2
RESTART=5
(none)
RESTART
RESTART=5
RSTIME=600
(none)