Foxy
The WINDOW command without an operand returns the current
setting.
Example:
Command:
Response:
WSTART=n
This command sets the window start time (n=0 to
359999 seconds) of the currently selected window.
Example:
Command:
Response:
The WSTART command without an operand returns the current
setting.
Example:
Command:
Response:
WEND=n
This command sets the window end time (n=0 to
359999 seconds) of the currently selected window.
Example:
Command:
Response:
The WEND command without an operand returns the current
setting.
Example:
Command:
Response:
UPLOAD
The unit responds to this command with the
current program settings, one line at a time. The UPLOAD
command must be sent for each line in the program until
NOTE=END is received from the unit. Settings are in the form of
commands allowing them to be sent back to the unit to restore a
program. All settings are sent, even if they are not relevant to
the way the program is currently set.
Example:
Command:
Response:
Command:
Response:
®
®
R1 and Foxy
R2 Fraction Collectors
Section 4 Serial Command Control
WINDOW
WINDOW=1
WSTART=36000
(none)
WSTART
WSTART=240
WEND=60
(none)
WEND
WEND=240
UPLOAD
PROGRAM=1
UPLOAD
DELAY=0
4-15