Cable For Non-Isco Detectors; Non-Isco Detector Cable, P/N 69-2134-171 - Teledyne Foxy R1 Instruction Manual

Hide thumbs Also See for Foxy R1:
Table of Contents

Advertisement

Table 1-8 Non-Isco Pump Cable, P/N 69-2134-170
Connector
1.6.3 Cable for non-Isco
Detectors
Table 1-9 Non-Isco Detector Cable, P/N 69-2134-171
Connector
Foxy
Pin
Color
Type
1
BLK
Input
2
BRN
Output
3
RED
Output
4
ORN
Common
5
YEL
Output
6
GRN
Not used
K
N/A
Shield
A cable is available to connect the fraction collector to a non-Isco
detector. Part number 69-2134-171 is a six-foot long cable with
one end terminated as a 8-pin mini-DIN connector. The other end
is stripped and tinned for connection to the external device.
Pin
Color
Type
1
BLK
Common
2
BRN
Input
3
RED
Not used
4
ORN
Output
5
YEL
Input
6
GRN
Input
7
BLU
Common
8
VIO
Input
K
N/A
Shield
®
®
R1 and Foxy
R2 Fraction Collectors
Section 1 Introduction
Description
Count: pump volume signal
Pump pause: stops pump between
tubes and at end of run
Valve output: control signal for
optional high flow diverter valve
Circuit common
+24 VDC Output: power for optional
high flow diverter valve
Earth ground
Description
Analog common for use with pin 8
Ext peak:
Event: pulse delivered at tube change
Inject: starts an inject cycle
Tube Advance
Circuit common
Analog detector signal: used for peak
detection
Earth ground
1-11