R2 Fraction Collectors
Section 2 Preparation for Use
Standard
Valve
Figure 2-1
Adjusting the height of the standard (left) and
high flow (right) diverter valves
2-2
Note
To perform the preliminary checkout you will need to power the
fraction collector. The fraction collector is shipped with either a
North American IEC320C13 to NEMA 5-15P power cord or a
European IEC320C13 to CEE7/VII power cord. If the cord
does not fit your AC mains power source, purchase a connec-
tor adapter or IEC320C13 power cord from a local vendor.
Do not install racks, tubing, or any accessories until the prelim-
inary checkout has been completed.
Should the fraction collector fail any preliminary checkout step,
contact the factory or your factory-authorized representative.
1. Set the fraction collector upright on a level surface.
2. Remove any wrapping or packing materials from the frac-
tion collector. The arm and diverter valve assembly should
move freely.
3. Loosen the diverter valve thumb screw(s) and raise the
diverter valve to its highest position (Figure 2-1). Tighten
the thumb screw(s) to secure the diverter valve.
4. Connect the fraction collector to AC mains power using the
supplied power cord. This places the fraction collector in
the Standby mode. The display should be off.
5. Touch the touch screen display to turn on the fraction col-
lector. The drop former will find the home position (left,
rear position) and the touch screen display backlight will
turn on. The display will briefly show the current rack con-
figuration and then the main menu.
CAUTION
High Flow
Valve