Cleaning The Check Valves - Thermo Scientific thermoscientific Vanquish Pumps C Operating Manual

Table of Contents

Advertisement

7.7.3

Cleaning the Check Valves

When
Parts and additional items required
Preparations
Follow these steps
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Dirty check valves
Once a year
Syringe (12 mL), filled with isopropanol
Ultrasonic bath
Remove the check valves that you want to clean. It is recommended to
clean all check valves of the pump head at the same time. To remove the
check valves, follow the related steps in
(} page 162)
or
Replacing the Outlet Check Valves (} page 164)
applicable.
1. In the direction of flow, fill the check valves with isopropanol, using
the syringe.
2. For 10 minutes, place the check valves in the ultrasonic bath.
3. In the direction of flow, rinse the check valves with isopropanol,
using the syringe.
4. Reinstall the check valves, following the related steps in
the Inlet Check Valves (} page 162)
Valves (} page 164)
5. In the Chromeleon software, consider updating the Predictive
Performance information for the check valves.
7
Replacing the Inlet Check Valves
or
Replacing the Outlet Check
as applicable.
Maintenance and Service
as
Replacing
Page 167