NOTICE
Observe the following:
Only use cleaning detergents that will not damage the surfaces of
the system.
Never use sharp tools or brushes for cleaning any surfaces.
Do not use sprays for cleaning.
Prevent cleaning detergent from entering the flow path.
Do not use excessively wetted cloth or wipes for cleaning. Prevent
any liquids from entering the functional components of the device.
Liquids can cause a short circuit when getting in contact with the
electronic components.
1. Turn off the power to the device and disconnect the power cord
from the power source.
1. Wipe the surfaces clean with a clean, dry, soft, lint-free cloth or
wipe. If necessary, slightly dampen the cloth or wipe with a solution
of lukewarm water and a suitable cleaning detergent.
2. Allow the cleaning detergent to react as recommended by the
manufacturer.
3. Wipe the cleaned surfaces with purified water to ensure that all
cleaning detergent residues have been removed.
4. Wipe the surfaces dry using a soft, lint-free cloth or wipe.
The Chromeleon software supports functions for estimating the lifetime
of consumables and for monitoring and recording service and
qualification information about the device. These functions are called
Predictive Performance. They allow you to schedule maintenance
procedures based on the actual operating and usage conditions of the
device.
On special wellness, service, and qualification panels, you can define
intervals for replacing components that are subject to wear or stress and
for service procedures or qualification procedures. In addition, you can
set limits to alert you before and when the replacement, service, or
qualification is due.