Follow these steps
5.6.2

Guiding Capillaries and Tubing Through the System

Tubing Chase, Tubing Guide, Tubing Bracket
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
To set up the additional flow connections and complete the installation,
follow these steps:
1. Connect the pump to the drain system (see
(} page 64)).
2. Connect the solvent lines (see
(} page 64)).
3. Set up the seal wash system (see
4. Connect the pump to the autosampler (see
and Autosampler
(} page 80)).
Flow connections between the modules of the Vanquish system are
guided through either the tubing chase in the devices or the guide holes
or capillary clips of the devices.
To guide certain tubes and lines (solvent tubing, wash liquid tubing,
detector waste line) from the top module to the bottom module in the
Vanquish system stack, the stackable modules have a tubing chase on
the inside right.
The tubing chase provides four tubing guides. Each guide can hold up to
three tubes or lines.
In each module, push the tube (or line) into the appropriate guide.
Figure 8: Tubing chase with tubing guides (left: view from inside, right:
view from top)
Guiding Liquids to Waste
Connecting the Solvent Lines
Seal Wash System
(} page 71)).
Connecting the Pump