Troubleshooting
8.5.3
Testing the Static Mixer for Permeability
When
Additional items required
Follow these steps
Page 212
TIP The information and references to the static mixer in this section
apply equally to the inline filter if installed.
If you observe pressure pulsation, inconstant pressure, or high
backpressure at the column and pump
•
Water as solvent
•
Tissue
Test the static mixer by following the steps in
Permeability
(} page 168).
Testing the Static Mixer for
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual