5
Installation
5.7.2

Setting Up the Seal Wash System

Parts and tools required
Page 72
Seal wash reservoir assembly, including reservoir, reservoir cap, cap
plugs, and retaining guide for the seal wash line
Seal wash line (silicone tubing) for connection to the seal wash
pump; the line is pre-installed to the seal wash outlet of the
metering device head
Seal wash line (silicone tubing) for connection to the seal wash
reservoir
NOTICE
If you want to use normal-phase compatible solvents and additives in a
Vanquish Core system, you must modify the VC-pump and VC-
autosampler with the components from the Normal-Phase (NP) kit. See
Normal-Phase Compatible Solvents and Additives
TIP To avoid the formation of gas bubbles in the solvent reservoirs, place
the reservoirs in the system stack on the same level or higher than the
pump. The setup description below assumes that the reservoirs are
placed in the solvent rack.
(} page 249).