Maintenance and Service

7.16.2 Returning the Slide-In Module

Preparation
Follow these steps
Page 192
3. Grasp the slide-in module by the leak tray and pull the module out
of the enclosure by approximately 10 cm.
Figure 63: Pulling out the slide-in module
NOTICE
The slide-in module can fall down when pulling it out of the enclosure
too far.
Pull out the slide-in module just far enough so that you can grasp it on
both sides from below.
4. Remove the slide-in module from the enclosure. The following steps
require a team effort:
a) Take the slide-in module on both sides from below.
b) Pull the slide-in module from the rails towards the front.
c) Place the slide-in module on a clean and stable surface.
If not yet done: Remove the slide-in module from the enclosure. See
Removing the Slide-In Module
1. Install the slide-in module to the enclosure of the replacement
device. Follow the steps for inserting the slide-in module in the
enclosure and tightening the captive screws in
Module
(} page 193).
2. Follow the instructions in
(} page 191).
Installing the Slide-In
Shipping the Device
(} page 190).
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)