2
Safety
2.4

Solvent and Additive Information

2.4.1

General Compatibility

Page 26
To protect optimal functionality of the Vanquish system, observe these
recommendations on the use of solvents and additives:
The system must be used with reversed-phase (RP) compatible
solvents and additives only.
Use only solvents and additives that are compatible with all parts in
the flow path.
TIP In a Vanquish Core system, normal-phase (NP) compatible solvents
and additives may be used if the VC-pumps and the VC-autosamplers are
modified with the components from the Normal-Phase (NP) kit. See
Normal-Phase Compatible Solvents and Additives
NOTICE
If the system flow path includes a Vanquish charged aerosol detector,
observe the specific solvent and additive information for this detector.
Refer to the Operating Manual for the Vanquish charged aerosol
detector.
(} page 249).