Digital I/O - Thermo Scientific thermoscientific Vanquish Pumps C Operating Manual

Table of Contents

Advertisement

13
Appendix

13.2 Digital I/O

Pin Assignment
Pin
Wire Color
1
Pink
2
Gray
Page 288
The digital I/O ports (Dig I/O) can be used to exchange digital signals
with external devices. Each port provides:
one digital input
one relay output
one bidirectional input/output
Figure 83: Digital I/O port
Pin
Description
1
Bidirectional input/output
2
Relay output — Relay_NC (Normally Closed contact)
3
Ground — GND
4
Digital input — Input
5
Relay output — Relay_COM
COM is the common contact for NO and NC. If the relay is not activated or if
the device is turned off, the connection is between COM and NC. If the relay
is activated, the connection is between COM and NO.
6
Relay output — Relay_NO (Normally Open contact)
The next table lists the functions assigned to the connector pins and the
color of the cable wire connected to each pin.
Signal Name
Signal Level
Input/output
Input (low active):
On: 0-0.4 V
Off: 2.2-5 V
Open Collector Output:
0-5 V, 0-2 mA
Pull-up resistor: 47 kΩ to 5 V
Relay output
Potential free
— Relay_NC
0-24 V, 0-100 mA
Remarks
Configure as either input or output.
Reference potential is ground. Note
the following:
The maximum input voltage at the
input must not exceed +5 V with
reference to ground.
The minimum input voltage must
not be lower than the ground
potential.
Opening contact
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)