Heat Therapy
Mattress Heater Control Panel
16
°C
>37 °C
<35 °C
1
Legend
(1)
Button for extending range of settings to
temperatures from 30° C to 38.5° C
(2)
Display of mattress set temperature
•
to the left of display: button for
decreasing set temperature
•
to the right of display: button for
increasing set temperature
(3)
Alarm silence button for silencing
intermittent audible alarm for 15 minutes;
•
Yellow LED in button illuminates while
intermittent audible alarm is silenced
(4)
Button for checking function of LEDs,
displays and audible alarm
(5)
Power switch
Figure 2. Babytherm Mattress Heater Control Panel
6
7
8 9 10
40 °C
±1 °C
Inop
°C
2
3
(6)
(7)
(8)
(9)
(10)
Babytherm 8000
Check
4
5
Display of actual (measured)
mattress temperature
Red LED
; illuminates in case
of power failure
Red LED 40° C; flashes when
measured mattress temperature
exceeds 40° C
Red LED ±1° C; flashes when
measured mattress temperature
deviates more than ±1° C from
set value
Red LED Inop; illuminates in
case of unit malfunction
!
DI0002