Karl Storz ENDOFLATOR 50 Instruction Manual page 89

Table of Contents

Advertisement

81
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Table 7 – PATIENT coupling PORT (according to IEC 60601-1-2)
Phenomenon
Basic EMC standard
ELECTROSTATIC DISCHARGE
IEC 61000-4-2
Conducted disturbances
IEC 61000-4-6
induced by RF fields
Table 8 – Signal input/output parts PORT (according to IEC 60601-1-2)
Phenomenon
Basic EMC standard
ELECTROSTATIC DISCHARGE
IEC 61000-4-2
Electrical fast transients/bursts
IEC 61000-4-4
Surges
IEC 61000-4-5
Line-to-ground
Conducted disturbances
IEC 61000-4-6
induced by RF fields
Electromagnetic
Compatibility (EMC)
information
Professional healthcare
facility environment
±8 kV Kontakt
±2 kV, ±4 kV, ±8 kV, ±15 kV air
3 V
0.15 MHz – 80 MHz
6 V in ISM bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
Professional healthcare
facility environment
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±1 kV
100 kHz repetition frequency
±2 kV
3 V
0.15 MHz – 80 MHz
6 V in ISM bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
Indicaciones sobre
compatibilidad
electromagnética (CEM)
IMMUNITY TEST LEVELS
HOME HEALTHCARE
ENVIRONMENT
Not applicable since released only
for professional healthcare facility
environment
IMMUNITY TEST LEVELS
HOME HEALTHCARE
ENVIRONMENT
Not applicable since released only
for professional healthcare facility
environment

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ui 500

Table of Contents