Karl Storz ENDOFLATOR 50 Instruction Manual page 86

Table of Contents

Advertisement

78
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Cables which have been shown to comply with IEC 60601‑1‑2:
Cable type
Shielded
PA
No
SCB Cable
Yes
Power Cord
No
031210-01
No
Guidance and manufacturer's declaration – electromagnetic emissions
The ENDOFLATOR
50 model UI 500 is intended for use in the electromagnetic environment specified below. The customer or user of the
®
ENDOFLATOR
50 model UI 500 should ensure that it is used in such an environment.
®
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Electromagnetic
Compatibility (EMC)
information
Length [m]
Ferrite
> 3
No
No
> 1
3
No
3
No
Compliance
Group 1
The ENDOFLATOR
®
Therefore, its RF emissions are very low and are not likely to cause interference in
nearby electronic equipment.
Class B
The ENDOFLATOR
®
domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Class A
– Professional healthcare facility environment –
Complies
Indicaciones sobre
compatibilidad
electromagnética (CEM)
Used for
Potential equalization
connecting SCB network, other SCB units
Power supply
Heatable insufflation tubing set
(patient tube heating)
Electromagnetic environment – guidance
50 model UI 500 uses RF energy only for its internal function.
50 model UI 500 is suitable for use in all establishments including

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ui 500

Table of Contents