Rectifying Faults
Fault
CO
concentration too low (deviation > 1%).
2
Possible cause
CO
bottle empty.
2
Inlet pressure or flow too low.
CO
valve closed or blocked.
2
Fault
CO
concentration too high (deviation < 1%).
2
Possible cause
CO
input pressure too high.
2
Fault
CO
value fluctuates, actual value is not constant.
2
Possible cause
Inlet pressure or flow too high.
Fault
The water supply for humidification is interrupted.
Possible cause
The water container is empty.
Air trapped between wet filter
and reservoir.
Valve defective.
Fault
humidity
The
parameter was deactivated automatically.
Possible cause
Transfer error of the humidity
sensor or temperature sensor.
Inlet pressure is too high.
Steam generator is defective.
Page 126 of 158
Faults in conjunction with the CO
Measures
Install new CO
bottle.
2
Increase inlet pressure or flow.
Contact your local INFORS HT branch.
Measures
Reduce input pressure.
Measures
Decrease inlet pressure or flow.
Faults in conjunction with the humidity parameter
Measures
Top up the water container.
Replace the wet filter with a dry one.
Contact your local INFORS HT branch.
Measures
Restart software (switch the device off and then on
again).
Reduce inlet pressure (max. 0.3 bar).
Contact your local INFORS HT representative.
parameter
2
Personnel
Operator
Operator
INFORS HT service tech-
nician or licenced dealer
Personnel
Operator
Personnel
Operator
Personnel
Operator
Operator
INFORS HT service tech-
nician or licenced dealer
Personnel
Operator
Operator
INFORS HT service tech-
nician or licenced dealer
20 August 2020
Need help?
Do you have a question about the Multitron and is the answer not in the manual?
Questions and answers
how to calibrate co2 on Model: Multitron Pro # I10102.1 and also purchase the solenoid valve # 52X06150GM
What is the meaning of Throw (fixed) and adjustable
In the Infors HT Multitron, "adjustable throw" means the shaking motion can be changed by selecting one of four eccentric throw positions: 12.5 mm, 19 mm, 25 mm, or 50 mm. This allows customization of the shaking intensity for different applications. A "fixed throw" would mean the shaking motion is set to a single, unchangeable value. The Multitron supports adjustable throw, and only trained personnel are allowed to make changes to it.
This answer is automatically generated