Multitron - Operating Manual
Options
Page 46 of 158
Prerequisites:
The level sensor is dismounted.
The water supply line is disconnected.
The lid of the vessel is mounted.
INFORMATION
The solenoid valve is closed in de-energized state (normaly closed,
NC). It is therefore not mandatory to fit a blind cap to the water
connection.
1.
Close the ports of the water connection and the level sensor
with a blind cap each:
a)
Ensure that on both ports a flat gasket is placed.
b)
Place the blind cap on the flat gasket.
c)
Fix the blind cap with the clamp.
2.
Close the water inlet line to the steam generators with the sup-
plied hose piece with rapid coupling.
3.
Before autoclaving ensure the following:
There is about 40 ml water in the vessel.
Level sensor is dismounted.
Hose piece with filter is open, overpressure can escape.
4.
Place the vessel in the autoclave.
5.
Autoclave the vessel according to the operating instructions of
the autoclave manufacturer.