Operation
Page 96 of 158
One-time change from phase 1 to phase 2
If the device is programmed in this way, the two different parameter
settings for the cultivating process change after the given period of
time. The device then runs using the settings for the second phase
until it is stopped manually by turning the parameters off.
Application examples:
Delayed start of the cultivation process
Induction of a protein expression
Slow down or stop the cultivation process after the given time
period
INFORMATION
To programme a one-time change from phase 1 to phase 2, the
timer for the second phase has to be switched off (
the timer for the second phase is switched on (
will be a cyclic change between the two phases (see the next ex-
ample).
Time 2
off
=
). If
Time 2
on
=
), there