Ac Input Wire; Ac Wire Input Connection - Xantrex XPR 10-600 Operating Manual

Xpr 6000 watt series programmable dc power supply
Table of Contents

Advertisement

AC Input Wire The manufacturer recommends the AC input wire specified in
be permanently connected to an approved AC distribution box with suitably rated
over-current protection. If you require a special cord, contact the manufacturer.
Table 2.2 AC Wire Specification for 6000 Watt units
AC Input Voltage Range
190–242Vac, 47–63Hz,
3-phase, 4 wire (standard)
342–500Vac, 47–63Hz,
3-phase, 4 wire (HV-Input)
AC Wire Input
See
Connection
To connect the AC input wires:
1. Ensure that the AC input cord is de-energized, and that the power switch on the
2. Strip approximately 4 in. (10 cm) from the jacket of the AC wire. Strip 0.55 in.
3. Undo the 2 screws for the AC wiring strain relief/cover on the rear panel.
4. Undo the strain relief screws. Insert the AC input cable through the strain relief
5. Insert the ground wire (green) 0.55 in. (14 mm) into the left-most terminal
6. Route the AC wires to the input terminal block by connecting the red, black, and
7. Reinstall the AC input strain relief/cover, routing wires inside the cover to
8. Connect the free end of the cable to the AC source, checking that the voltage is
9. Energize the AC input.
Release 1.1
Figure 2.5, on page
34.
front of the power supply is OFF.
(14 mm) at the end of each wire.
Remove the cover.
until the outer cable jacket is flush with the inside of the strain relief. Tighten the
strain relief cable clamp screws.
location, and tighten securely. (The safety ground wire may alternatively be
connected to the chassis ground stud next to the terminal block, using a suitably
sized ring terminal).
white wires to the remaining 3 cable clamp connectors. There is no set order for
connecting the wires. Any of the 3-phase wires can be connected to any of the 3
line input connectors. To connect each wire, loosen the terminal screw, insert the
stripped wire 0.55 in. (14mm) into the terminal, and tighten the screw securely.
prevent pinching.
within the approved input range for the supply.
Wire
4 x 10 AWG (3 wire plus safety ground), stranded
°
copper, 60
C minimum, 300V, 0.800 in. maximum
cable diameter, rated for 25A.
4 x 14 AWG (3 wire plus safety ground), stranded
°
copper, 60
C minimum, 600V, 0.800 in. maximum
cable diameter, rated for 13A.
Installation
AC Input Power
Table
2.2. This must
33