Xantrex XPR 10-600 Operating Manual page 76

Xpr 6000 watt series programmable dc power supply
Table of Contents

Advertisement

Specifications and Characteristics
Output Performance Specifications
Fall Time
For a programmed 95% to 5% step in output voltage.
Load Condition
No Load
Full Load
Time Delay From Power On Until Output Stable
5 s maximum
(Within regulation envelope)
Time Delay From Output Enable Until Output Stable
2 s maximum
(Within regulation envelope)
Output Hold-Up Time – Power Off
Minimum 4 ms (at full load)
Output Hold-Up Time – Source Interruption
Minimum 4 ms with output deviation less than 5% of maximum output voltage after
source interruption.
Transient Response Time
Time to recover within 0.75% of rated output of previous level after step change in
load current between 50% and 100%.
Mode
Voltage Mode
Mode Crossover
Maximum deviation as a percentage of rated output voltage.
CV – CC Overshoot
Peak–Peak and RMS Noise Bandwidth Limits
The frequency range for Peak to Peak measurements is 10 Hz–20 MHz.
The frequency range for RMS measurements is 10 Hz–100 kHz.
1.
74
Time (Max)
3 s
50 ms
Time
3 ms
1%
Time for the output voltage to recover within 0.75% of rated output of its previous level
after a step change in load current of up to 50% - 100% and 100% to 50% of rated
output
1
Operating Manual for XPR Series Power Supply