completely dry. Wet or moist hands may cause electric shock.
4)
Avoid placing objects in the front and rear of the instrument.
5)
Avoid installing the instrument in a dusty environment. Excessive dust may cause malfunction or
damage to the instrument.
6)
Avoid installation near heat sources. This can cause fire.
7)
Avoid installation near sources of water or damp locations. This can cause electrical shock, fire or
instrument malfunction.
8)
Do not install near sources of flammable or corrosive gas. If there is a gas leak, do not touch the
power plug but open a window to circulate fresh air. Sparks from the power plug can cause fire and
explosions.
9)
Do not disassemble or modify the instrument in any way. This can result in fire, electrical shock or
malfunction, and also voids your warranty.

3. Precautions and warnings regarding instrument installation

1)
This is a precision instrument. Do not install in a location exposed to direct sunlight.
2)
Install the instrument on a flat, solid surface that is flat and does not move.
3)
When installing the instrument, make sure at least 15 cm separate the instrument from the nearest
wall.
4)
Take caution not to damage the cooling fan mesh (located on the front-bottom) while installing.
5)
When connecting the PC, network hub and ExiPrep
powered off. Connecting components while the power is on may cause damage to the instrument.
6)
Verify the network cable integrity when connecting between the PC, network hub and ExiPrep
Dx. Unstable network connections may result in invalid experimental data from data communication
errors.

4. Precautions and warnings regarding instrument operation

1)
Dust off the power plug and insert the plug so that the connection is firm and does not wiggle.
Incomplete electrical contacts may cause fire.
2)
Operate the instrument in an ambient temperature range of 15 ~30 . Excessive exposure to heat
may affect the instrument and yield inexact results.
3)
Operate the instrument within the recommended humidity range (20~80%, no condensation). Humid
conditions may cause corrosion or malfunction.
4)
Do not place any objects next to or behind the instrument. The instrument may malfunction.
5)
This instrument contains precision machined parts. Do not drop or severely agitate the instrument.
This can break the instrument and compromise the safety of the product.
6)
When not using the instrument for a long period of time, turn the instrument off and unplug from the
wall outlet. Overheating and fire may occur.
7)
Take caution not to damage the cooling fan mesh located on the front-bottom of the instrument. If